qLabs
®
ElectroMeter
Users Manual
For Health Care Professional Use Only
REF Q-3 Plus

Questions and Answers

  • W
    whitechristopherSep 3, 2025
    What does it mean if the MICROPOINT qLabs ElectroMeter displays 'External QC failed'?
    • M
      Melissa BurkeSep 3, 2025
      If your MICROPOINT Medical Equipment displays 'External QC failed', ensure that the test strip and control are within their expiry period. Also, verify that you have entered the correct test strip Codechip number and control Codechip number.
  • D
    Debra AlvaradoSep 2, 2025
    What to do if the MICROPOINT qLabs ElectroMeter displays 'Internal QC failed'?
    • Z
      zthomasSep 2, 2025
      If your MICROPOINT Medical Equipment shows that the internal QC failed, use an unexpired test strip and enter the correct test strip Codechip number.
  • B
    Brandon CarrSep 1, 2025
    What to do if the MICROPOINT qLabs ElectroMeter Medical Equipment says the sample is insufficient?
    • J
      Jasmine PhillipsSep 1, 2025
      If your MICROPOINT Medical Equipment indicates that the added sample is insufficient, retest with a new test strip, making sure to add enough blood for the test.
  • L
    Lonnie DickersonAug 31, 2025
    Why does the MICROPOINT Medical Equipment say the sample may not be suitable for qLabs testing?
    • W
      William BlakeAug 31, 2025
      If the MICROPOINT Medical Equipment indicates that the sample may not be suitable for qLabs testing, retest with a new test strip. If the problem continues, contact your local dealer and consider using a laboratory method for the test.
  • C
    Christina HallAug 30, 2025
    What to do if the MICROPOINT qLabs ElectroMeter says the sample adding has timed out?
    • J
      jimwolfeAug 30, 2025
      Ensure that the sample adding is completed before the end of the countdown on your MICROPOINT Medical Equipment.
  • B
    Brian MckinneyAug 29, 2025
    What should I do if the MICROPOINT qLabs ElectroMeter Medical Equipment says the sample is added too early?
    • J
      John JeffersonAug 29, 2025
      If the MICROPOINT Medical Equipment indicates that the sample was added too early, use a new test strip and add the sample only after the meter prompts you to do so.
  • C
    Christine HernandezAug 28, 2025
    What to do if the MICROPOINT Medical Equipment shows a print error?
    • M
      Michelle HaasAug 28, 2025
      If your MICROPOINT Medical Equipment displays a print error, ensure that the meter is properly communicating with the base. Also, check that the printing paper at the base is installed correctly and that there is sufficient paper.
  • M
    Mrs. Hannah Smith MDAug 27, 2025
    What to do if the MICROPOINT qLabs ElectroMeter displays 'Codechip reading error'?
    • F
      Fred HallAug 27, 2025
      If your MICROPOINT Medical Equipment shows a Codechip reading error, unplug the Codechip and reinsert it, ensuring it is fully inserted. If the problem continues, contact technical support.
  • N
    Nicole HayesAug 26, 2025
    What to do if heating has timed out on MICROPOINT qLabs ElectroMeter Medical Equipment?
    • S
      Stephanie BarrAug 27, 2025
      If the heating process times out on your MICROPOINT Medical Equipment, adjust the ambient temperature to between 10 °C and 35 °C. Turn off the meter for 5 minutes, then turn it back on and restart the test. If the problem continues, contact technical staff.
  • M
    Mason TaylorAug 25, 2025
    What should I do if the MICROPOINT Medical Equipment is showing a Bluetooth error?
    • B
      Brian FordAug 26, 2025
      If your MICROPOINT Medical Equipment shows a Bluetooth error, turn the meter off and reboot it. If the problem persists, please contact technical support.

Summary

Introduction

Before You Start

Instructions to read the manual before using the device.

Intended Use

Purpose of the qLabs monitoring system for quantitative measurement.

Test Principle

How the qLabs ElectroMeter detects coagulation and determines results.

Packaging

Information on how the qLabs ElectroMeter is packed and storage advice.

Product Overview

Operation Summary

Before Testing

Precautions, Limitations and Warnings

Care of Your qLabs ElectroMeter

Guidelines for handling, storing, and maintaining the qLabs ElectroMeter.

Patient Health Status

Importance of considering patient health for accurate test interpretation.

Performing a Test

Instructions and precautions for operating the qLabs ElectroMeter during a test.

Collecting a Fingerstick Blood Sample

Steps for properly collecting a fingerstick blood sample for testing.

Collect fresh venous whole blood samples

Procedures for collecting venous blood samples for the qLabs system.

Electromagnetic Compatibility

Information on EMC compliance and potential interference.

Power On;Off

Power On

Steps to turn on the qLabs ElectroMeter and initial setup.

Power Off

Procedures for turning off the qLabs ElectroMeter and auto-shutdown.

Settings

Enter Settings Screen

How to access the main settings menu on the qLabs ElectroMeter.

Language

How to select the desired language for the qLabs ElectroMeter interface.

Bluetooth

Instructions for enabling, disabling, and connecting the qLabs ElectroMeter via Bluetooth.

Test Parameter

How to configure test parameters like Sample No., Codechip, and Reference Range.

System Parameter

Configuration of system parameters like Date/Time, Beeper, and Screen Rotation.

About the Device

Viewing device information such as version and system log.

Operator ID

Setting up and managing operator identification and password on the meter.

Sample Test

Input Sample No.

Entering the sample number for a test, manually or automatically.

Insert a Test Strip

How to correctly insert a test strip into the qLabs ElectroMeter.

Input Test Strip Information

Entering the Codechip number for the inserted test strip.

Install Codechip

Installing the test strip Codechip into the meter's chip slot.

Heating

The meter enters a heating state before sample addition.

Add Sample

Instructions for adding the blood sample to the test strip.

Perform a Test

Initiating and monitoring the test after sample addition.

Test Results

Displaying and interpreting the PT/INR and APTT test results.

Quality Control

Insert a Test Strip

Inserting a test strip for quality control testing.

Input Test Strip Information

Entering Codechip information for the QC test strip.

Install Test Strip Codechip

Installing the Codechip for the QC test strip into the meter.

Input QC Information

Entering the Codechip number for the QC liquid.

Install QC Liquid Codechip

Installing the Codechip for the QC liquid into the meter.

Heating

The meter enters the heating state for QC testing.

Add Sample

Adding QC liquid sample to the test strip.

QC Test

Performing the quality control test with the QC liquid.

QC Test Results

Displaying and interpreting the results of the quality control test.

Results

Result Channel

Accessing the results query screen to view test and QC results.

Test Result

Viewing detailed information of individual test results.

QC Results

Viewing detailed information of individual QC results.

Maintenance

Care and Cleaning of Your qLabs ElectroMeter

Instructions and frequency for cleaning the qLabs ElectroMeter.

Recommended cleaning;disinfecting solutions

Recommended cleaning solutions and cloths for the qLabs ElectroMeter.

Cleaning and disinfecting the meter housing

Step-by-step guide for cleaning and disinfecting the meter's exterior.

Precautions for Lithium-ion Battery

Safety precautions and charging advice for the device's lithium-ion battery.

Servicing

Information on servicing and repair of the qLabs ElectroMeter.

Troubleshooting

Symbols

Operating Condition and Product Specifications

Operating Condition

Environmental conditions required for operating the qLabs ElectroMeter.

Product Specifications

Technical specifications of the qLabs ElectroMeter, including processor, display, and power.

Special storage conditions and methods

Warranty

Limited Warranty

Terms and conditions of the limited two-year warranty for the qLabs ElectroMeter.

MICROPOINT qLabs ElectroMeter Specifications

General IconGeneral
Sample Volume10 μL
Operating Temperature10°C to 40°C
ConnectivityUSB
Dimensions120 x 65 x 25 mm
Weight150 g
DisplayLCD
Sample TypeWhole blood
Measuring Range (PT/INR)0.8 - 8.0 INR
Storage Temperature (Test Strips)2°C to 30°C
Measurement Range0.8 - 8.0 INR
Input ImpedanceNot Applicable
Storage TemperatureNot Applicable
Power Supply2 x AAA batteries