31
(27) COND AVG TIMES
Conduction test times 1-99.
(28) SEQ. STAT
General: Statistics is performed once the test has completed.
MultiDUT MUL.DU: When the test has completed, multiple DUTs will be included in the
statistics.
(29) AF.TSET TIME
Total test time for fixture current box; the test results within this time are aggregated and
averaged. 0 means single test.
(30) SAVE MODE
AUTO: Saves when the exit button on the setting screen is pressed.
MANUAL: The save button must be pressed for saving.
(31) TEST ALL M.D FAIL
Fail CONTINUE: MultiDUT test is used with continuous testing under all test completed mode;
when there are DUTs that fail, continue testing until test has completed.
All Fail STOP: MultiDUT test is used with continuous testing under all test completed mode;
stops when all DUTs fail.
(32) A FIX. 0.1-4A OUT.
ALL ON: All channels output current.
ONE ON: Fixed single channel outputs current.
(33) OS MISS WIRE
ON: Enables the OS misplacement detection function.
OFF: Disables the OS misplacement detection function.
(34) CK.CBL.Exist (HV))
ON: Gives a warning when DUT is removed before high voltage test is completed; the DUT
must have a short-circuit breaker.
OFF: Disables the warning function.