30
(15) Mul.DUT OPT
4: The maximum number of DUTs that can be set for MultiDUT is 4 sets.
14: The maximum number of DUTs that can be set for MultiDUT is 14 sets (expansion box
required for use with REMOTE signal output).
(16) SF.Stat MODE
General: Statistics is performed once the test has completed.
Empty: After DUT has been removed when test has completed, statistics is performed once the
short-circuit breaker is in an empty state.
MultiDUT MUL.DU: When the test has completed, multiple DUTs will be included in the
statistics.
(17) SET PULSE TIME
The pulse hold time 1-1000s when pulse mode is used for the REMOTE signal output.
(18) Mul.DUT OUTSGL
PASS: Test result is PASS output signal.
FAIL: Test result is FAIL output signal.
(19) Seq.Result Mode
Individual MERGE: Output result for each step individually.
Merge ALOND: Merge (OR) output results of each step.
(20) PC LINK CMD
ON: Used with official connection software.
OFF: Not used with official connection software.
(21) TEST COUNTER
Gives a warning when the number of tests reaches the number of times set; 0 disables this
function, 0-999999.
(22) COND.ZERO. RANGE
When the offset value exceeded the value set, give up offset 1-45.
(23) R Offs. Delay
When component R is testing delay offset 3000-60000.
(24) AF.DELAY TIME
When the fixture current box is testing, delay 1-20000mS after outputting current.
(25) Default Setting
ON: Execute reset to default setting.
OFF: Do not execute reset to default setting.
(26) OS CHECK TIMES
If the number of short-circuit test results failed is the same as the number of times set, determine
the result as fail 1-1000.