29
STATIC: Remote control is continuous output until the next test. When you are unsure of the PLC
reception speed, please set it as continuous mode, and note that the output signals all have a pull-up
resistor of 4.7K when sent to PLC input.
PULSE: When the remote control has a pulse output of 200ms. Users can set the pulse width according
to their needs.
(7) STEP TRIGGER
Selects the step start mode. There are three modes. KEY, AUTO, and EXTERNAL.
(8) TEST FILE
Choose whether to use SINGLE or MULTI-STEP test.
SINGLE: Current file test (which is the {Current file} displayed on the main screen).
MULTI-STEP: Step file selection test.
AUTO: Use step file test when there are step files, and use current file test if there are none.
(9) STEP TEST FAIL
Processing method to use when the step file test failed. Available selections include BACK FIRST,
REPEAT FIRST, BREAK TEST and TEST ALL.
(10) STEP DISP ALL
Whether to display the steps when using step file test.
ON: Yes
OFF: No
(11) FAIL LOCK
Lock the keys when test failed. Test signal must be sent remotely in order to unlock.
(12) TEST Statistic
Whether the keep statistics of the test results.
(13) Test Data Up Load
Function to select whether to upload the test results.
OFF: Do not upload the test results.
USB STORE: Generates a test result file on the USB flash drive.
TO PC USB: Transmits the test result to the computer terminal through the RS232 interface.
(14) CONNECT .DET
Tests the step files one by one and ends the test when PASS is matched. The test result is PASS. If the
test results of all step files failed, the test result is FAIL. When this function is enabled, the function
becomes invalid when "[9] Step test fail”.