Printer (Optional)
Operating the Printer
Whenthesterilizerispluggedin,theprinteris automatically powered up and initialized.
No user intervention or setup is required.
Printer Tape Description
The printer will print the following information for each program cycle:
•CycleNumber
•Identication:alineisprintedsothesterilizerI.D.canberecordedontheprintertape
• Operator:alineisprintedsotheoperatorssignaturecanberecordedontheprintertape.
•Date:linesareprintedforthemonth,day,year,hour,andminutesothedateandtimecanbe
recorded by the operator on the printer tape.
•“BEGINselectedCYCLE”:toindicatethebeginningofthecycleselectedbytheoperator.
•Summaryofselectedcyclesetpoints.
Oncethecyclestarts,theprinterwillprintthewords“FILLINGCHAMBER”toshowthatthesterilizerisling
with water.
OncethesterilizerbeginstheHeatingPhaseofthesterilizationcycle,theword“HEATING”isprintedand
theprinterwillprintthechambertemperature,pressure,andelapsedtimein30sec.incrementsuntilthe
heating phase is completed.
Whenthesterilizerentersthesterilizationphase,theword“STERILIZING”isprintedandtheprinterwillprint
thechambertemperature,pressure,andelapsedtimein30sec.incrementsuntilthesterilizationphaseis
completed.
Whenthesterilizerhascompletedthesterilizationphaseoftheprogramcycle,theprinterwillprintthewords
“VENTINGCHAMBER”toshowthatthesteampressureisbeingexhaustedfromthechamber.
Whenthesterilizerentersthedryingphase,theword“DRYING”isprintedandtheprinterwillprintthe
elapsedtimein5minuteincrementsstartingwith0:00andthewords“DRYINGSTART”.Theprinter
continuestoprinttheelapsedtimein5minuteincrementsuntilthedryingphaseiscompleted.Thenal
recordforthedryingphasewillincludethewords“DRYINGCOMPLETE”.Intheeventthedryingtimeis
programmedtoatimethatisn’tdivisibleby5,thenalprintedrecordforthedryingphasewillreectthe
actualprogrammeddryingtimein1minuteincrements,e.g.aprogrammeddrytimeof12minuteswillhave
5,10,and12minutesprintedontheprintertape.
Whenthesterilizerhascompletedthedryingphaseofthesterilizationcycle,theprinterwillprintasummary
of the sterilization cycle with the duration of each phase of the cycle and the Total Cycle Time. Following the
summarytheprinterwillprint“CYCLECOMPLETE”.
[NOTE: If drying cycle is aborted, “DRYING COMPLETE” and “CYCLE COMPLETE” will not print].