EasyManua.ls Logo

Midmark M9 - Example of Typical Printout of a Program Cycle

Default Icon
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English - 36
003-2707-99
© Midmark Corporation 2015 TP202 20-42-FO-00014 Rev A1 C2169
Example of Typical Printout of a Program Cycle
Midmark M9 - v1.0.3
Total Cycles: 10
SterilizerID
Operator
mm / dd / yyyy hh : mm
BEGINUNWRAPPEDCYCLE
Temp: 270 Degrees F
Time: 3 Minutes
Dry: 30 Minutes
FILLINGCHAMBER
HEATING
mm:ss Degrees PSI
0:00 233.8 F 0.0
0:30 231.3 F 1.9
2:30 265.7 F 25.0
3:00 271.3 F 28.8
STERILIZING
mm:ss Degrees PSI
0:00 271.8 F 28.9
0:30 272.4 29.5
2:30 272.7 29.3
3:00 272.6 29.3
Min 271.8 F 28.7
Max 273.2 29.9
VENTINGCHAMBER
DRYING
MM:SS
0:00 Drying Start
5:00
25:00
30:00 Drying Complete
FILLING: 3:22
HEATING: 8:54
STERILIZING: 3:00
VENTING: 2:16
DRYING: 30:00
TOTALCYCLE 00:47:32
CYCLECOMPLETE
- - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - -
- - / - - / - - - - - - : - -
Line provided for
sterilizer identication
Total # of cycles
run on unit
Line provided for
operator signature
Space provided for
recording date & time
Selected Cycle
Indicates Filling
phase initiated
Sterilization temperature,
time, and dry time settings
During Heat-Up phase, the
printer records chamber
temperature & pressure in
30-second increments
During Sterilization phase,
the printer records chamber
temperature & pressure in
30-second increments
Summary of
Sterilization phase
Indicates Venting
phase initiated
During Drying phase,
the printer records time in
5-minute increments
These lines will show the
total dry time completed
These lines will not print
if Dry Cycle is aborted
before it completes
Cycle summary