English - 30
003-2707-99
© Midmark Corporation 2015 TP202 20-42-FO-00014 Rev A1 C2169
Weekly Maintenance
Tray Rack
Tray Plate
Trays
• Clean Chamber / Trays (including Rack and Plate)
Note
Step A is not necessary if the sterilizer is connected to a direct-to-drain thermal reduction system.
Equipment Alert
Failure to change water may result in sterilizer malfunction. Do not use bleaching
agents or any abrasive materials / substances in chamber (i.e. bleach, steel wool, wire
brush, scouring powder, etc.). Failure to comply may result in damage to the chamber and/or other
components.
Note
Everysevendays,theautoclavewillautomaticallydisplaythePERFORMWEEKLYMAINTENANCE
message.Ifpowerisdisconnected,thecycleofweeklymessageswillbereset.
A) Disconnect the upper portion of the reservoir drain tube from the panel clips, bend it downward,
and drain the reservoir water into a suitable container, e.g. a bucket, and dispose of the water.
B) Remove the trays, tray rack, and tray plate from the sterilizer.
(Refer to the following page for instructions on removing / installing the tray rack and tray plate).
C) Wash trays, rack, plate, and inside of chamber with mild soap or Speed-Clean and distilled water or
water that meets the referenced water purity specifications.
D) Refill reservoir with distilled water or water that meets the referenced water purity specifications.