English - 40
003-2707-99
© Midmark Corporation 2015 TP202 20-42-FO-00014 Rev A1 C2169
Informational Messages
The chart below lists the informational messages that may appear during operation.
Error Messages
Thechartbelowandonthefollowingpages,liststheerrormessagesthatmayappearduringoperation.
Message Possible Cause Solution
INITIALIZING SYSTEM
Unit power cord was just plugged
in - standard informational
message.
Normal operation will occur after a
4 second pause.
TOTAL CYCLES
XXX M9*, VX.X.X
(*will display model
number,M9,M11)
Unit power cord was just plugged
in - standard informational
message.
Normal operation will occur after this
message.
PERFORM WEEKLY
MAINTENANCE
This message is displayed every
7,14,and21daysaftertheunit
is plugged into a power source to
prompt the operator to perform
weekly maintenance described in
this manuals.
Performweeklymaintenance.
The message will automatically clear
afterthenextcycleiscompleted.
PERFORM MONTHLY
MAINTENANCE
This message is displayed every
28 days to prompt operator to
perform monthly maintenance
described in this manual.
Performmonthlymaintenance.
The message will automatically clear
afterthenextcycleiscompleted.
Caution
If an error occurs more than once, stop using the sterilizer. Record the message or error
code, unplug unit, then call an authorized service representative. (see “Calling for Service”)
If an error message includes: “Items Not Sterile”, the items in sterilizer shall not be considered sterile;
they must be run through a successful sterilization cycle.
Message Possible
Cause
Solution
C010: POWER UP MODE
SYSTEM PWR LOSS
ITEMS NOT STERILE
PUSH STOP TO RESTART
Unit had loss of power
during cycle.
PressSTOP button to restart.
This chart continues on the following page...