Error Messages - continued
Message Possible Cause Solution
CO60: POWER UP MODE
SYSTEM HARDWARE
Powerwasinterrupted
brieyoraninternal
glitch.
Unplug unit power cord for 1 minute and then plug
backin.Ifproblempersists,contactanauthorized
servicerepresentative.(see“CallingForService”)
C102: FILL MODE
STOP PRESSED
STOPbuttonwaspressed
during cycle.
PressSTOPbuttontorestart.
C103: THROUGH C105
HEATUP, STERILIZE OR
VENT MODE
STOP PRESSED
STOPbuttonwaspressed
during cycle.
Waitbriey(uptooneminute)whilechamberpressure
/temperaturedissipates.PressSTOPbuttontoreturn
to select cycle mode where a new cycle may be
initiated.
C106: DOOR MODE
STOP PRESSED
STOPbuttonwaspressed
during cycle.
PressSTOPbuttontorestart.
C232: FILL MODE
WATER LOW
Water level in
reservoir is too low.
Rellwaterreservoirwithdistilledwater,orwater
thatmeetsthereferencedwaterpurityspecications.
Waitbriey(uptooneminute).PressSTOPbuttonto
return to select cycle mode where a new cycle may be
initiated.
Fill/Ventlter(inbottom
of chamber) is clogged.
CleanFill/Ventlter.
(see“MonthlyMaintenance”)
C326: DOOR MODE DOOR
CLOSED
Door latch safety switch is
still making contact after
door motor operated.
Open door.
C382: FILL
DOOR OPEN
Sterilizer detects that door
safety switch contacts
opened.
Close the sterilizer door.
(Cycle will continue where left off.)
C383: HEATUP MODE
DOOR OPEN
Sterilizer detects that door
safety switch contacts
opened.
Waitbriey(1minute)whilechamberpressure/tem-
peraturedissipates.PressSTOPbuttontoreturnto
selectcyclemode.Initiateanewcycle.
C384: STERILIZE MODE
DOOR OPEN
Sterilizer detects that door
safety switch contacts
opened.
Unplug unit power cord for 1 minute and then plug
backin.Ifproblempersists,contactanauthorized
servicerepresentative.(see“CallingForService”)
C533 through C633:
STEAM TEMP LOW
or
STEAM TEMP HARDWARE
or
PRESSURE LOW
Sterilizer detects that
temperature and / or
pressure is outside the
limits for normal
operation.
Unplug unit power cord for 1 minute and then plug
backin.Ifproblempersists,contactanauthorized
servicerepresentative.(see“CallingForService”)
This chart continues on the following page...