VetPro DC Installation/Service Manual, 00-02-1606, Revision R01
106
Figure 99
31. Re-attach the two ground cables that were removed in (Step 6.)
Figure 100
32. Connect the black and white connectors that were disconnected in
(Step 5.)
Figure 101
33. To ensure correct operation and to prevent damage to the yoke cable
assembly perform the following steps:
33.1) Point the tubehead down.
33.2) Check that the large Phillips head screw is at the 9 o'clock
position and that there is a natural loop in the cable harness.
There should be a
natural loop in the
cable harness
Large Phillips head
screw at the 9
o'clock position