Milli-Q® Reference System 67
Replacing the Quantum® Cartridge, Continued
Rinsing
(continued)
Step
Action
Diagram
4
Dispense water for about
10 minutes.
This flushes out any trapped air
in the System.
This also rinses off the
purification media located in the
Q-Gard® Pack and the
Quantum® Cartridge.
5
Leave the System in READY
Mode when finished.