66 Milli-Q® Reference System
Replacing the Quantum® Cartridge, Continued
The Quantum® Cartridge, when newly installed, needs to be rinsed. This ensures
optimal water quality.
Locate the clear tubing and the
barbed fitting from the System
accessories bag.
Screw the barbed fitting onto
the POD Unit.
Do not use any white tape on
the threads of the barbed fitting.
An O-ring is located inside the
POD Unit.
Push one end of the clear tubing
onto the end of the barbed
fitting.
Place the other end of the clear
The System must be in READY
Mode.
Push the plunger down on the
POD Unit.
In a few minutes, water should
dispense from the
POD Unit.
Continued on next page