66 Milli-Q® Reference System
Replacing the Quantum® Cartridge, Continued
Rinsing
The Quantum® Cartridge, when newly installed, needs to be rinsed. This ensures
optimal water quality.
Step
Action
Diagram
1
Locate the clear tubing and the
barbed fitting from the System
accessories bag.
Screw the barbed fitting onto
the POD Unit.
NOTE:
Do not use any white tape on
the threads of the barbed fitting.
An O-ring is located inside the
POD Unit.
Push one end of the clear tubing
onto the end of the barbed
fitting.
Place the other end of the clear
tubing into a sink.
2
The System must be in READY
Mode.
3
Push the plunger down on the
POD Unit.
In a few minutes, water should
dispense from the
POD Unit.
Continued on next page