37
Mahr GmbH, Millimar C1208/C1216/C1240
6. Press the START key again to adopt the
flashing nominal value as the current setting.
The display then returns to the catalog of
functions and parameters item “MASTER”.
If at this point either the MASTER or DATA
key is pressed instead of START, the dis-
play returns to the numerical input stage (cf.
point 4). The indicated nominal value can be
changed again if required.
7. If no further settings are to be made in the
catalog of functions and parameters, press
the START key. The standard display ele-
ments appear.
If further settings are necessary, navigate
to the appropriate point of the catalog of
functions and parameters using the ESC,
MASTER, DATA and MENU keys and carry
out the settings as required.
8.3 Entering the nominal master value
for a two-point master mea sure-
ment
The dimensions of master workpieces usually
deviate only fractionally from the required nomi-
nal size. The actual size of the master workpiece
must be entered in the catalog of functions and
parameters prior to starting master measure-
ment.
1. Press the MENU key. The catalog of func-
tions and parameters item “FEATURE (1)” is
displayed.
2. Press the MENU key again.
3. Use the DATA and MASTER keys to select
MSTR.MAX”.
4. Press the MENU key again. “MSTR.MAX
is displayed along with a numerical value
that will vary depending on the measuring
unit that has been selected for the display:
0.3 mm or 300 µm or 0.011811 inch.
5. Change the indicated numerical value to
match the nominal value of the larger of the
two masters (see section “Changing numeri-
cal values for parameter settings” in Chapter
0).
6. Once the nominal value has been set as re-
quired, press the START key. This terminates
the process and the set nominal value flash-
es.