INSTALLATION
- 17 -
4-12 INSTALLATION AND RINSING OF THE FINAL FILTER
Obtain a Final Filter for example Millipak® Express 20 Filter.
Remove the Tygon Tubing and the Barbed Fitting from the POU Dispenser.
Screw the Final Filter onto the end of the POU Dispenser. The Final Filter should be turned until it is
hand tight (A). Do not over tighten the Final Filter.
Do not use white tape on the threads of the Final
Filter. The POU Dispenser has an O-ring inside
which provides a watertight seal.
It is highly recommended not to put Tygon Tubing or
any other type of tubing on the end of the Final
Filter. This can compromise the Product Water
quality (bacteria can grow in the Tygon Tubing).
Press the Main Button to put the system into DISPENSING mode (B).
Locate the air vent valve on the top side of the Final FIlter. Open this slowly but do not remove it from
the Millipak Filter (C). Allow any air on the clear side of the Millipak Filter to be vented out.
Dispense about 1 Litre of water. Make sure all air is purged out.
Press the Main Button. The system stops dispensing water.
Leave the system in PRE OPERATE mode (see Section 5-4 Operating Modes).
The Water System is now ready for use.
A
B
C
ATTENTION
ATTENTION