MAINTENANCE
- 33 -
6-4 HOW TO REPLACE THE FINAL FILTER
The Final Filter is normally replaced when the SimpliPak Cartridge is replaced or at an earlier time if it
becomes clogged. A clogged Final Filter can reduce the Product Water flow rate.
Make sure the SimpliPak Cartridge has been
hydrated overnight.
Remove the used Final Filter.
Screw the new Final Filter onto the end of the POU Dispenser. The Final Filter should be turned until it
is hand tight (A). Do not over tighten the Final Filter.
Do not use white tape on the threads of the Final
Filter. The POU Dispenser has an O-ring inside which
provides a watertight seal.
Press the Main Button to put the system into DISPENSING mode (B).
For the SimFilter, dispense about 0.5 Litre of water. For the BioPak® UltraFilter, refer to the
instructions shipped with the filter.
Press the Main Button. The system stops dispensing water.
Leave the system in PRE OPERATE mode.
The Water System is now ready for use.