USING THE ICW-3000 SYSTEM
- 26 -
5-6 HOW TO UNDERSTAND ICW-3000 SYSTEM MESSAGES
PACK ALARM
The system will prompt you to change the SimpliPak Cartridge
using a red Pack Alarm icon. The Display will show the red
Pack Alarm blinking. The SimpliPak Cartridge is changed due to
either the amount of time it has been used or from the amount of
water that has passed through it.
When the Resistivity display is blinking, the red Pack Alarm will
also be blinking. This indicates that the SimpliPak Cartridge
should be replaced. This message is shown when the Product
resistivity is less than the Resistivity Setpoint. The Resistivity
Setpoint can be seen in menu option C02
When the red Pack Alarm is displayed as a steady icon, the
SimpliPak Cartridge is not installed correctly or not installed at
all. If the SimpliPak Cartridge has been reinstalled and the
Alarm is still displayed, then contact Millipore.
UV
LAMP ALARM
The system will prompt you to change the UV Lamp using a
red UV Lamp Alarm icon. The message is shown when the UV
Timer displays 0 days. The UV Timer can be viewed in menu
option C04. The replacement of the UV Lamp involves the
installation of a new UV Lamp and a manual reset of the UV
Timer.
When the red UV Lamp Alarm is displayed as a steady icon, the
UV Lamp is not installed correctly or not installed at all. If the
UV Lamp has been reinstalled and the Alarm is still displayed,
then contact Millipore.