MAINTENANCE
- 37 -
6-6 HOW TO VIEW OR RESET THE UV LAMP TIMER (C03)
The UV Lamp Timer should be reset only after the UV Lamp has been replaced (see Section 6-5 How to
Replace the UV Lamp). The UV Lamp Timer displays the time left until the UV Lamp needs to be replaced.
The Display will show the red UV Lamp Alarm icon blinking when the Timer reaches 0 days. This message
is displayed until the UV Lamp is replaced and the UV Lamp Timer is reset.
H
OW TO VIEW THE REMAINING DAYS ON THE UV LAMP TIMER
1. Press the Main and
“+“ Buttons together
to enter the menu.
The Display will
show C01.
2. Press the Main
Button 3 times. The
Display will show
C03 and the days
left on the UV
Timer.
3. To exit the menu,
press and hold the
Main Button for
2 seconds.
To display the next
menu option, press
the Main Button
once.