MAINTENANCE
- 36 -
I
NSTALLING THE NEW UV LAMP
Ensure that you use the gloves supplied with the
UV replacement kit. Plug the electrical cable to
the new UV Lamp (G).
Carefully insert the UV Lamp into the UV
housing (H).
Attach the UV housing with the Velcro belt (I).
Install the SimpliPak Cartridge (J).
Install the Final Filter.
Close the front cover.
Plug the electrical cord to power ON the system.
Reset the UV Timer. See Section 6-6 How to
View or Reset the UV Lamp Timer (C03).