4-14 HOW TO SET THE RESISTIVITY SETPOINT (C02)
The Resistivity Setpoint is used to inform you when the Product resistivity is low. When the resistivity is
below the setpoint, the Resistivity display flashes and the red Pack Alarm is blinking (see Section 5-6 How
to Understand ICW-3000 System Messages). The factory default resistivity value is set to
15 M.cm @25 C.
1. Press the Main and “+“
Buttons together to enter the
menu. The Display shows
C01.
2. Press the Main Button once.
The Display shows C02 and
the Resistivity Setpoint
value.
3. Press the “+” or “-“ Button to
adjust the Resistivity Setpoint
from 1.0 M.cm @25 C
to18.0 M.cm @25 C.
If Conductivity Units are chosen in C01, then the Setpoint can be adjusted from 0.999 S/cm @25 C to
0.055 S/cm @25 C.
NOTE: The Conductivity Setpoint display needs to be multiplied by 0.001 to get the real value. For
example, if the Display reads “055” S/cm @25 C, then you would multiply 055 x 0.001 = 0.055. Thus,
the real Conductivity Setpoint reading is 0.055 S/cm @25 C.
4. To exit the menu, press and
hold the Main Button for 2
seconds.
To display the next menu
option, press the Main Button
once.