14 - 2 BeneVision N Series Patient Monitor Operator’s Manual
• If you need to measure SpO
2
 by both the MPM module and the SpO
2
 module, select the same type of 
SpO
2
. Otherwise, the SpO
2
 module will be disabled. For example, if an MPM module configured with 
the Nellcor SpO
2
 and an SpO
2
 module configured with the Mindray SpO
2
 are simultaneously 
applied, the SpO
2
 module will be automatically disabled.
• The SpO
2
 extension cable should be compatible with the SpO
2 
connectors. For example, you can 
only connect the Mindray SpO
2
 extension cable to the Mindray SpO
2
 connectors.
• A functional tester or SpO
2
 simulator can be used to determine the pulse rate accuracy.
• A functional tester or SpO
2
 simulator cannot be used to assess the SpO
2
 accuracy.
14.2 SpO
2
 Safety Information
• When a trend toward patient deoxygenation is indicated, analyze the blood samples with a 
laboratory co-oximeter to completely understand the patient’s condition.
• Do not use SpO
2
 sensors during magnetic resonance imaging (MRI). Induced current could 
potentially causes burns. The sensor may affect the MRI image, and the MRI unit may affect the 
accuracy of the oximetry measurements. 
• Prolonged continuous monitoring may increase the risk of undesirable changes in skin 
characteristics, such as irritation, reddening, blistering or burns. Inspect the sensor site every two 
hours and move the sensor if the skin quality changes. Change the application site every four hours. 
For neonates, or patients with poor peripheral blood circulation or sensitive skin, inspect the sensor 
site more frequently.
• If the sensor is too tight because the application site is too large or becomes too large due to edema, 
excessive pressure for prolonged periods may result in venous congestion distal from the 
application site, leading to interstitial edema and tissue ischemia. 
• When patients are undergoing photodynamic therapy they may be sensitive to light sources. Pulse 
oximetry may be used only under careful clinical supervision for short time periods to minimize 
interference with photodynamic therapy. 
• Setting alarm limits to extreme values may cause the alarm system to become ineffective. For 
example, high oxygen levels may predispose a premature infant to retrolental fibroplasia. If this is a 
consideration, do not set the high alarm limit to 100%, which is equivalent to switching off the 
alarm. 
• SpO
2
 is empirically calibrated in healthy adult volunteers with normal levels of carboxyhemoglobin 
(COHb) and methemoglobin (MetHb). 
• Change the application site or replace the sensor and/or patient cable when a persistent SpO2 Low 
Signal Quality message is displayed on the equipment. These messages may indicate that patient 
monitoring time is exhausted on the patient cable or sensor.
• Replace the cable or sensor when a “SpO2 Sensor Off”, “SpO2 No Sensor”, or “SpO2 Low Signal 
Quality” message is consistently displayed while monitoring consecutive patients after completing 
troubleshooting steps listed in this manual. 
• Variation in measurements may be profound and may be affected by sampling technique as well as 
the patient's physiological conditions. Any results exhibiting inconsistency with the patient’s clinical 
status should be repeated and/or supplemented with additional test data. Blood samples should be 
analyzed by laboratory instruments prior to clinical decision making to completely understand the 
patient’s condition.
• Use only SpO
2
 sensors specified in this manual. Follow the SpO
2
 sensor’s instructions for use and 
adhere to all warnings and cautions.