39 - 14 BeneVision N Series Patient Monitor Operator’s Manual
39.13 The Authorization Setup Settings
Parameters On/Off
/
Selects what parameter s can be monitored.
Parameter 
Output Setup
Baud Rate Off Configures DIAP protocol parameters to realize 
communications between the monitor and third 
party devices.
Parity Mode None
Data Bits 8
Stop Bits 1
Menu Item Default Setting Description
Section Menu Item Default Setting Description
/ Retention Time 20 sec Selects timeout period of the MLDAP password 
for accessing the Maintenance menu, alarm 
settings and arrhythmia settings. 
If there is no operation after the specified 
timeout period is reached, you need to re-enter 
the password.
Maintenance User Maintenance Local Password Selects the password for accessing the monitor’s 
Maintenance menu.
•
Local Password: the monitor’s password for 
accessing the 
Maintenance menu is required.
•
User Password: the user name and password 
saved in the MLDAP server are required.
Modify Local Password / Changes the monitor’s password for accessing 
the 
Maintenance menu.
Others Alarm Setup No Password Selects the password for changing alarm settings.
•
No Password: changing alarm settings is not 
password protected.
•
Local Password: changing alarm switch, 
alarm limit, and alarm priority is password 
protected. The monitor’s password for 
changing alarm settings is required.
•
User Password: changing alarm switch, alarm 
limit, and alarm priority is password 
protected. The user name and password 
saved in the MLDAP server are required.
Arrhythmia No Password selects the password for changing arrhythmia 
settings.
•
No Password: changing arrhythmia settings 
is not password protected.
•
Local Password: changing arrhythmia switch, 
alarm priority, and arrhythmia threshold is 
password protected. The monitor’s password 
for changing arrhythmia settings is required.
•
User Password: changing arrhythmia switch, 
alarm priority, and arrhythmia threshold is 
password protected. The user name and 
password saved in the MLDAP server are 
required.
Modify Local Password / Changes the monitor’s password for accessing 
alarm settings and arrhythmia settings.