D - 12 BeneVision N Series Patient Monitor Operator’s Manual
D.2.17 NMT Technical Alarm Messages
D.2.18 rSO
2 
Technical Alarm Messages
Alarm message Default priority
Indication on 
alarm reset
Cause and solution
NMT No Cable Low A Check that NMT patient cable is properly 
connected to the NMT module.
NMT No Sensor Low A Check that NMT sensor is properly connected to 
the NMT patient cable. If the alarm persists, 
replace the sensor.
NMT Stimulation 
Electrode Off
Low A Check that NMT sensor is properly connected to 
the NMT patient cable. If the alarm persists, check 
the application of electrodes.
NMT Sensor Error Low C Contact your service personnel.
Alarm message Default priority
Indication on 
alarm reset
Cause and solution
rSO2-1 No Sensor Low A rSO
2
 sensor is detached, or the sensor cable is 
disconnected from the pre-amplifier. Re-connect 
the sensor to the pre-amplifier.
rSO2-2 No Sensor
rSO2-1’ No Sensor
rSO2-2’ No Sensor
rSO2-1 Excess Light Low C The sensor is detached or ambient light is too 
strong. Re-attach the sensor or reduce the level of 
ambient light.
rSO2-2 Excess Light
rSO2-1’ Excess Light
rSO2-2’ Excess Light
rSO2-1 Low Signal Quality Low C The signal acquired by the sensor was unstable or 
weak due to power supply noise. Check the 
sensor connection. If the alarm persists, contact 
your service personnel.
rSO2-2 Low Signal Quality
rSO2-1’ Low Signal Quality
rSO2-2’ Low Signal Quality
rSO2 -1/rSO2-2/rSO2-1’/
rSO2-2’ No Preamplifier
Low A Properly connect the pre-amplifier.
rSO2 -1/rSO2-2/rSO2-1’/
rSO2: Change Sensor
Low C The sensor type mismatches patient category, 
the reusable sensor cable fails, the sensor reaches 
the lifetime, or the sensor type mismatches the 
monitor and patient category setting. Verify 
patient category and use sensors of correct 
patient category.
rSO2 -1/rSO2-2/rSO2-1’/
rSO2-2’ Interference
Low C There is noise interference (such as 
electrosurgical unit). Check   for any possible 
sources of signal noise. 
rSO2-1 Auto Baseline Prompt / The monitor automatically sets the rSO
2
 baseline.
rSO2-2 Auto Baseline
rSO2-1’ Auto Baseline
rSO2-2’ Auto Baseline
Disconnect/ Reconnect 
rSO2
Low A Remove the rSO
2
 module and reconnect it.