BeneVision N Series Patient Monitor Operator’s Manual D - 15
Note: XX refers to the department name, YY refers to the room number, and ZZ refers to the bed number.
D.2.24 Technical Alarm Messages Related to Telemetry Monitors and BP10 Modules
LAN1 IP Address Conflict Low C Wired network LAN1 IP network conflicts. Check 
the network settings.
Fail To Get WLAN IP 
Address
Low C Unable to automatically obtain the wireless 
network IP address. Check the network settings.
Fail To Get LAN1 IP 
Address
Low C Unable to automatically obtain the wired 
network LAN1 IP address. Check the network 
settings.
Alarm message Default priority
Indication on 
alarm reset
Cause and solution
Alarm message Default priority
Indication on 
alarm reset
Cause and solution
Telemetry Disconnected High B The TM80 is power-off powered off, the monitor 
or the TM80 is not connected to the Wi-Fi 
network, or out of the network coverage.
1. Power on the TM80.
2. Connect the monitor and the TM80 to the 
network.
3. Move the TM80 in the network coverage.
The exchanger network connected by the TM80 
and the monitor does not support multicast data 
transfer. 
Contact your service personnel.
Telemetry NIBP 
Disconnected
High B The BP10 is power- off, or out of the MPAN 
coverage.
1. Power on the BP10.
2. Move the BP10 in the MPAN coverage area.
Telemetry Error High C An error occurred to the TM80.
Restart the TM80. If the problem still persists, 
replace with a known good TM80.
Telemetry Low Battery Med C The TM80 battery charge is low.
Replace with a known good battery.
Telemetry Battery 
Depleted
High C The TM80 battery charge is critically low.
Replace with known good batteries.
Telemetry Battery 
Maintenance Required
Med C The TM80 lithium-ion battery may reach its life. 
Replace with known good batteries.
Telemetry Battery Error Med C The TM80 lithium-ion battery communication 
encounters an error. Replace with known good 
batteries.
Telemetry Battery Type 
Error
Med C The TM80 battery contacts may fail. Replace with 
known good batteries.
Telemetry NIBP Error High C An error occurred to the BP10. Restart the BP10. If 
the problem persists, replace with a known good 
BP10.
Telemetry NIBP Low 
Battery
Med C The BP10 battery charge is low. Replace with a 
known good battery.
Telemetry NIBP Battery 
Depleted
High C The BP10 battery charge is critically low. Replace 
with a known good battery.
Telemetry NIBP Battery 
Maintenance Required
Med C The BP10 lithium-ion battery is may reach its 
lifetime. Replace with a known good battery.