18 - 6 BeneVision N Series Patient Monitor Operator’s Manual
18.8 C.O. Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or 
accessories, check the table below before requesting for services. If the problem persists, contact your service 
personnel.
• For the physiological and technical alarm messages, see DAlarm Messages.
Problem Solution
Do not see C.O. numeric area on the main 
screen
1. Check that the C.O. is set to display in the Screen Setup menu. For 
more information, see 
39.9.5The Other Tab.
2. Check that if the C.O. parameter switch is enabled. If not, enable the
C.O. measurement. For more information, see 
3.11.1Switching On or Off 
a Parameter
.
3.Check that the patient type is adult.
4. Check the connection of C.O. cable, thermodilution catheter and TI 
sensor.
C.O. value is inaccurate 1. Check that the thermodilution catheter is positioned properly.
2. Check that the computational constant is proper for current injectate 
temperature, injectate volume and injectate probe type.
3. Inject solution rapidly and smoothly.
4. Finish injection within four to five seconds.
5. Inject more volume, or inject colder solution.
6. Check that the height and weight of patient is properly configured.
7. If 
Auto TI is switched off, check that the entered temperature is 
correct.
C.O. measurement fails 1. Inject more volume, or inject colder solution. Make sure that the 
injectate temperature is at least 10°C colder than the patient blood 
temperature.
2. Finish injection within four to five seconds.
3. Check the connection of C.O. cable, thermodilution catheter and TI 
sensor.