BeneVision N Series Patient Monitor Operator’s Manual 39 - 19
39.17.10 The Information Security Tab
39.17.11 The MLDAP Tab
Section Menu Item Default Setting Description
Data+Waveforms Server Address / Inputs the name or IP address for the server 
receiving the realtime data and waveform.
Destination IP 0.0.0.0
Port 0 /
Send Data Off
Data Interval 30 sec
Send Waveforms Off
Alarms Server Address / Inputs the name or IP address for the server 
receiving the alarm data.
Destination IP 0.0.0.0
Port 0 /
Send Alarm Off
Menu Item Default Setting Description
Encryption Connection Type Only Private 
Encryption
• Only Private Encryption: Mindray private 
encryption is used to encrypt the transmitted 
data. You cannot connect devices supporting 
SSL (secure sockets layer) encryption.
•
SSL Encryption Priority: for devices 
supporting SSL encryption, SSL encryption is 
used when connecting the devices. For 
devices not supporting SSL encryption, 
private encryption is used when connecting 
the devices.
Broadcast Patient Demographics On
On: when viewing other patients, device location 
and patient information of remote devices are 
displayed in the remote device list.
•
Off: patient information does not display in 
the remote device list.
Menu Item Default Setting Description
Server Address / Inputs the name or IP address for the MLDAP 
server.
IP Address 0.0.0.0
Port 0 /
Network Test / Tests whether the monitor is properly connected 
with the MLDAP server.