BeneVision N Series Patient Monitor Operator’s Manual 16 - 1
16 Monitoring Noninvasive Blood Pressure (NIBP)
16.1 NIBP Introduction
The monitor uses the oscillometric method for measuring the non-invasive blood pressure (NIBP). NIBP 
measurement is based on the principle that pulsatile blood flow through an artery creates oscillations of the 
arterial wall. The oscillometric device uses a blood pressure cuff to sense these oscillations that appear as tiny 
pulsations in cuff pressure. The Oscillometric devices measure the amplitude of pressure changes in the 
occluding cuff as the cuff deflates from above systolic pressure. The amplitude suddenly increases as the pulse 
breaks through the occlusion in the artery. As the cuff pressure decreases further, the pulsations increase in 
amplitude, reach a maximum (which approximates to the mean pressure), and then diminish. The oscillometric 
method measures the mean pressure and determines the systolic and diastolic pressures.
The NIBP module is integrated into the MPM module. NIBP monitoring is intended for adult, pediatric, and 
neonatal patients.
• Blood pressure measurements determined with this device are equivalent to those obtained by a 
trained observer using the cuff/stethoscope auscultatory method or an intra-arterial blood pressure 
measurement device, within the limits prescribed by the American National Standard: manual, 
electronic, or automated sphygmomanometers. 
• NIBP measurement can be performed during electro-surgery and discharge of defibrillator.
(1) NIBP cuff connector