A.3 Environmental Specifications
• The monitor may not meet the performance specifications if stored or used outside the specified 
temperature and humidity ranges. If the performance of the equipment is degraded due to aging or 
environmental conditions, contact your service personnel.
• When the monitor and related products have differing environmental specifications, the effective 
range for the combined products is that range which is common to the specifications for all 
products.
• The environmental specification of unspecified parameter modules are the same as those of the 
main unit.
Components Item Operating Condition Storage Condition
Main Unit Temperature (ºC) 0 to 40 
(0 to 35 for N17 configured 
with iView module)
-20 to 60
Relative humidity 
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 427.5 to 805.5 120 to 805.5
Microstream CO
2
 
module
Temperature (ºC) 0 to 40 -20 to 60
Relative humidity 
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 430 to 790 430 to 790
Sidestream CO
2
 
module
Temperature (ºC) 5 to 40 -20 to 60
Relative humidity 
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 430 to 790 430 to 790
Mainstream CO
2
 
module
Temperature (ºC) 0 to 40 -20 to 60
Relative humidity 
(noncondensing) (%)
10 to 90 10 to 90
Barometric (mmHg) 427.5 to 805.5 400 to 805.5
AG module Temperature (ºC) 10 to 40 -20 to 60
Relative humidity 
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 525 to 805.5 525 to 805.5
RM module Temperature (ºC) 5 to 40 -20 to 60
Relative humidity 
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 427.5 to 805.5 120 to 805.5
ICG module Temperature (ºC) 10 to 40 0 to 50
Relative humidity 
(noncondensing) (%)
15 to 95 15 to 95
Barometric (mmHg) 427.5 to 805.5 120 to 805.5