C-10 Description of Self Test Items
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace TRA board if the test result is FAIL.
C.1.18 DSP FPGA and TRA Interconnection Test (Data
Interface)
Top test item
DSP FPGA and TRA FPGA communication self test (control interface)
Test content
Test whether XCVER data interface between TRA FPGA on the main board and DSP FPGA works
well.
Analysis to test failure
The drive goes wrong if the test result appears Error.
XCVER data interface on the connection of TRA FPGA-DSP FPGA goes wrong if the test result is
FAIL.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace TRA board if the test result is FAIL.
C.1.19 PC Module and DSP FPGA Interconnection Test
Top test item
N/A
Test content
Test whether PCIe interface communication between PC module and DSP FPGA works well.
Analysis to test failure
Check whether Windows device manager recognizes Ultrasound Backbone Device. If the drive is
not installed properly, it cannot be recognized.
The drive goes wrong if the test result appears Error.
PC module and DSP FPGA has communication error if the test result is FAIL.
Suggestion to failure test
It is recommended to restore the device (OS+doppler) if the drive is not recognized.
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace engine board if the test result is FAIL.
C.1.20 DSP FPGA DDRIII Test
Top test item
PC module and DSP FPGA interconnection test

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-70 Pro and is the answer not in the manual?

Mindray DC-70 Pro Specifications

General IconGeneral
Touch ScreenYes
Touch PanelYes
Transducer Ports4
Battery BackupYes
3D/4D ImagingYes
Operating SystemWindows-based
Power Supply100-240V AC, 50/60 Hz
Storage500GB HDD
ConnectivityUSB, HDMI
ApplicationsCardiology, Obstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
PortsUSB, Ethernet, HDMI
Display Size19 inch