Description of Self Test Items C-11
Test content
Perform read and write tests for entire space of four types DDRs of DSP FPGA. The program
displays the test results of DSP buffer DDR (transmitting DSP processing result data), SCAN
buffer DDR (transmitting scan control frame), IQ buffer DDR (transmitting IQ data) and Gather
buffer DDR (collecting data).
Analysis to test failure
The drive goes wrong if the test result appears Error.
The connection error between FPGA and plug-in DDR occurs if the test result is FAIL.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace engine board if the test result is FAIL.
C.1.21 ATGC Function Test
Top test item
TRA AFE test mode
Test content
Control a channel to transmit 2 V waveform. Collect the waveform that AFE receives as setting
ATGC to max, min and medium value. Decide whether the amplitude of waveform changes as
ATGC increases according to the analysis on RMS value.
Analysis to test failure
The drive goes wrong if the test result appears Error.
The circuit of ATGC gain adjustment goes wrong if the test result is FAIL.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace engine board if the test result is FAIL.
C.1.22 MF FPGA and PHV ARM Interconnection Test
Top test item
Communication self test between PC Module and MF FPGA
Test content
Test whether UART communication between DSP FPGA and PHV ARM works well. Send the
order through PHV serial port drive, and re-read the data via the port.
Analysis to test failure
The drive goes wrong if the test result appears Error.
UART interconnection between FPGA and PHV ARM goes wrong if the test fails.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to change PHV power supply board if the test result is FAIL.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-70 Pro and is the answer not in the manual?

Mindray DC-70 Pro Specifications

General IconGeneral
Touch ScreenYes
Touch PanelYes
Transducer Ports4
Battery BackupYes
3D/4D ImagingYes
Operating SystemWindows-based
Power Supply100-240V AC, 50/60 Hz
Storage500GB HDD
ConnectivityUSB, HDMI
ApplicationsCardiology, Obstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
PortsUSB, Ethernet, HDMI
Display Size19 inch

Related product manuals