C-20 Description of Self Test Items
If the flicking works well, the program passes, and then exits the test.If the flicking fails to work,
click “Touch Screen is Bad”.
Analysis to test failure
The touch screen does not work if the test result is FAIL.
Suggestion to failure test
Confirm the drive of the touch screen is installed properly.
Then confirm whether connecting material between the touch screen and main unit is fixed well.
It is recommended to replace touch screen.
C.1.39 Mini CW Board ID Test
Top test item
DSP FPGA and TRA FPGA communication self test (control interface)
Test content
The program reads the signal in the place to decide whether mini CW board is in the place. Then,
read CW board’s ID. SPI interface communication between TRA board and mini CW board works
well if it appears non-0 and non-F.
Analysis to test failure
The drive goes wrong if the test result appears Error.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-70 Pro and is the answer not in the manual?

Mindray DC-70 Pro Specifications

General IconGeneral
Touch ScreenYes
Touch PanelYes
Transducer Ports4
Battery BackupYes
3D/4D ImagingYes
Operating SystemWindows-based
Power Supply100-240V AC, 50/60 Hz
Storage500GB HDD
ConnectivityUSB, HDMI
ApplicationsCardiology, Obstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
PortsUSB, Ethernet, HDMI
Display Size19 inch

Related product manuals