12-2 Care and Maintenance
No.
Item
Frequency
Method
3.
Trackball cleaning
1 time/month
Ditto
4.
Control Panel/minor panel cleaning
1 time/month
Ditto
5.
Probe cleaning (head of the probe)
Every time after
use
Ditto
6.
Probe cable and connector cover
cleaning
1 time/month
Ditto
7.
Holder cleaning
1 time/month
Ditto
8.
Cover cleaning
1 time/month
Ditto
9.
Peripherals cleaning
1 time/month
See Chapter 12.2.2
10.
Probe appearance check
1 time/day
See Chapter 12.3.1
11.
Check for power supply cable, power
supply plug.
1 time/month
See Chapter 12.3.1
12.
Battery check
2 times/year
See Chapter 12.3.1
13.
Check for peripherals and optional
functions
1 time/year
See Chapter 12.3.3
14.
Mechanical safety check
1 time/year
See Chapter 12.3.4
15.
Check for electrical safety
Every two years
See Appendix A