C-6 Description of Self Test Items
Analysis to test failure
The drive goes wrong if the test result appears Error.
SPI between TRA FPGA and AFE has communication error if the test result is FAIL.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to change TRA board if the test result is FAIL.
C.1.14 Z0403 TRA AFE Digital Interface Test
Top test item
TRA AFE SPI test
Test content
Enter system test mode, write the data to TRA AFE, input delay RAM, re-read the data from
delay-channel memory to locate the channel and AFE clip.
Analysis to test failure
The drive goes wrong if the test result appears Error.
AFE clip goes wrong if the test result is FAIL.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error, which indicates the logic
error of DSP FPGA occurs. It is necessary to restore the device (OS+doppler) if Error re-appears.
It is recommended to change TRA board if the test result is FAIL.
C.1.15 Z0404 TRA Transmission and Reception Function
Test
Top test item
TRA AFE test mode
Test content
Test the coherence of 64 channels in transmitting and receiving: one channel transmits 2 V 1 M
PHV1 waveform each time, and analyzes whether the waveform is PHV1, and compares the
signal-to-noise ratio of this channel with others’.
Analysis to test failure
The drive goes wrong if the test result appears Error.
The waveform that the channel transmits is incorrect if the information “Open circuit of transmitting
and receiving channel: XXX” appears.
The transmission of this channel affects other channels if the information “Short circuit of
transmitting and receiving channel” appears.
Suggestion to failure test
Restart the device to perform the self test if the test result appears Error. It is necessary to restore
the device (OS+doppler) if Error re-appears.
It is recommended to replace TRA board if the test result is FAIL.

Table of Contents

Other manuals for Mindray DC-80A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80A and is the answer not in the manual?

Mindray DC-80A Specifications

General IconGeneral
BrandMindray
ModelDC-80A
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals