Image Optimization 5-19
Map
Description
This function is a combination of several image parameters, which
indicates the display effect of color image.
Operation Select among the maps through the [Map] item in the soft menu or menu.
You can also adjust through the color bar: move the cursor to the color bar
and press <Set> on the control panel to adjust.
The system provides 21 different maps to be selected among, where the V
group provides 11 ordinary maps and the VV group provides 10 2-D maps.
Impacts The function is available in real-time imaging, freeze or cine review status.
Color IP
Description
Color IP is a combination of several image processing parameters, which is
used for a fast image optimization. The IP combination number is displayed
in the image parameter area in the upper left corner of the screen.
The Color IP combination parameters include persistence and smooth.
Operation Select among the maps through the [Color IP] item in the soft menu or
menu.
Or adjust it in the image parameter area.
The system provides 8 groups of IP combinations, and the specific value of
each parameter can be preset.
WF (Wall Filter)
Description
It filters out low-frequency signals to provide effective information, and this
function is used to adjust the filtered frequency. The real-time value is
displayed in the image parameter area in the upper left corner of the
screen.
Operation Select through the [WF] item in the soft menu or menu.
Or adjust it in the image parameter area.
There are 8 levels of wall filter function, adjust according to the actual
situation and the probe.
Impacts Flow signals may be missing.
Priority
Description
This function is used to set levels of the flow display, to display the
grayscale signal or color signal.
Operation Select the value through the [Priority] item in the soft menu or menu.
The adjusting range is 0-100%.
Effects The higher the value, color signals are prior to be displayed; while the
lower the value, grayscale signals are prior to be displayed.
iTouch
Description
To optimize image parameters as per the current tissue characteristics for
a better image effect.

Table of Contents

Other manuals for Mindray M7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray M7 and is the answer not in the manual?

Mindray M7 Specifications

General IconGeneral
TypePortable Ultrasound System
Display15-inch LCD
Imaging Modes2D, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Cardiac, Vascular, OB/GYN, Small Parts, Musculoskeletal, Urology, Pediatric
BatteryRechargeable Lithium-ion
ConnectivityUSB, Ethernet, DICOM
Dimensions35 cm x 36 cm x 9 cm
Transducer Ports2 active transducer ports