Image Optimization 5-23 
Adjust through [Trace Smooth] on the menu. 
There are 4 levels of smooth effect provided, the bigger the value, the higher the 
smooth processing. 
This function is used to set the sensitivity of tracing in the spectrum. 
Adjust through [Trace Sensitivity] on the menu. 
There are 5 levels of sensitivity adjustment, the bigger the value the higher the 
sensitivity. 
In real-time scanning, the results displayed are derived from the calculation of 
the latest cardiac cycle. 
In the freeze and cine status, the results displayed are calculated from the 
current selected area. 
 
Invert 
This function is used to set how the spectrum is displayed. 
Turn the function on or off using the [Invert] control. 
Select “Auto Invert” in the [Setup] -> [System] -> [Image], so the spectrum can 
automatically invert when the color flow is steered to a certain angle to 
accommodate the operator’s desire to distinguish the flow direction. 
 
Speed 
This function is used to set the scanning speed of PW mode imaging. 
Adjust through [Speed (mm/s)] on the menu. 
There are 6 levels of scan speed available. 
Changing the speed makes it easier to identify the cardiac cycles and to detect 
more details. 
 
T/F Res. 
This function is used to create a balance between time resolution and spatial 
resolution. 
Adjust through [T/F Res.] on the menu. 
There are 4 levels of T/F Res. values available. 
 
Wall Filter 
It filters out low-velocity signals to provide effective information, and this function 
is used to adjust the filtered frequency. The real-time value is displayed in the 
image parameter area in the top-left corner of the screen. 
Adjust through [WF] on the menu. 
7 levels of wall filter function are provided. 
Signals of low-velocity flow may be missing.