4-16 
The difference is within ±3 mm when 0, 50 or 200 mmHg is 
set for NIBP accuracy test. 
 
There is no leakage with NIBP, or the manual leakage test 
result does not exceed 6 mmHg/min. 
 
CO
2
 test 
The difference is within  ±0.3% when gas level 5% is set for 
CO2 accuracy test. 
 
Temp test 
The accuracy of temperature on monitor is within  ±0.2Ԩ 
with THP79JU Infrared ear thermometer when the target 
temperatures are 37.0Ԩ and 42.0Ԩ. 
 
The accuracy of temperature on monitor is within  ±0.1Ԩ 
with Covidien Genius 2 Tethered Tympanic Thermometer 
when the target temperature is 37.0Ԩ. 
The accuracy of temperature on monitor is within  ±0.2Ԩ 
with Covidien Genius 2 Tethered Tympanic Thermometer 
when the target temperature is 42.0Ԩ. 
 
The accuracy of temperature on monitor is within  ±0.1Ԩ 
with Exergen TemporalScanner Thermometer when the 
target temperatures are 16.0Ԩ and 43.0Ԩ. 
 
The displayed Temp value on the monitor is within 37 ± 0.1 
ºC with SmarTemp
TM
 module. 
 
Nurse Call Reply Performance Test 
When an alarm is reported on the monitor, a nurse call is sent 
out through the cable. 
 
Barcode Scanner Test 
The indicator turns green and scanned characters are 
displayed on the monitor. 
 
Electric Safety Tests 
Refer to . All the electrical safety tests should be passed.     
Touchscreen Calibration 
The touchscreen is calibrated successfully.     
Recorder check 
The recorder can print correctly and the printout is clear.     
Set the recorder to some problems such as out of paper, etc. 
the monitor gives corresponding prompt messages. After the 
problem is removed, the recorder is able to work correctly. 
 
Automatic alarm recording for each parameter functions 
correctly when parameter alarms occur. 
 
Battery check