6. Operating | 6.2 - Configuration and adjustment via the keypad | 108
6.2.6 Displaying adjustment data for dead load and Max
Normally, all device settings are documented/printed out upon
commissioning.
If a faulty device is replaced by a new one and no new adjustment is
performed, the values for the dead load and the max. capacity (Max) in mV/V
must be obtained.
1. Press and hold for 2–3 seconds.
SEtuP appears on the display.
2. Press "OK".
Cd 000 appears on the display (Cd flashes).
3. Press to change the parameter number (000 flashes).
4. Press (3×) to select parameter number 003.
5. Press "OK".
dEAdLo appears on the display.
6. Press "OK".
byLoAd appears on the display.
7. Press to select byMV-V.
8. Press "OK".
The mV/V value for the dead load appears on the display.
9. Note the displayed value.
10. Press "OK".
Cd 004 appears on the display.
11. Press "OK".
SPAn appears on the display.
12. Press "OK".
byMV-V appears on the display.