6. Operating | 6.2 - Configuration and adjustment via the keypad | 122
Condition SPM bit Description
b66 = 1 X66 = 1 Marker bit 3 set, markers are reset after
power on.
6.2.11 Configuring digital outputs
The desired function for output1 to output3 is configured by selecting a
signal from the list.
The output then adopts the corresponding state, see example.
SPM addresses for the digital outputs, see ChapterSystem data.
Selection list for output functions
Function SPM bit Description
b00 = 0 X00 = 0 Digital input 1: inactive
b01 = 0 X01 = 0 Digital input 2: inactive
b02 = 0 X02 = 0 Digital input 3: inactive
b16 = 0 X16 = 0 Limit signal 1: inactive
b17 = 0 X17 = 0 Limit signal 2: inactive
b18 = 0 X18 = 0 Limit signal 3: inactive
b32 = 0 X32 = 0 General error in weighing point: inactive
(no error)
b33 = 0 X33 = 0 Weight above Max: inactive
b34 = 0 X34 = 0 Weight above Max plus overload value: in
active
b35 = 0 X35 = 0 Weight not below zero
b36 = 0 X36 = 0 Weight not within ¼d around zero
b37 = 0 X37 = 0 Weight not in zeroset range
b38 = 0 X38 = 0 Standstill inactive
b39 = 0 X39 = 0 Weight not below zero or above Max
b48 = 0 X48 = 0 For internal use only.
b49 = 0 X49 = 0 For internal use only.
b50 = 0 X50 = 0 Set after power on (= power fail): inactive
b56 = 0 X56 = 0 Analog test not started.

Table of Contents