6. Operating | 6.2 - Configuration and adjustment via the keypad | 126
6.2.12 Configuring the analog output
Parameter table
Parameter Selection Description
[Off] Analog output is unused.
[Transparent] Value output in D30.
Instrument is controlled via PLC.
[Gross] Gross weight output in D8.
[Net] Net weight output in D9; only if tared, other
wise gross weight is output.
[Mode]
[Selected] Gross/net value output on the display (D11) de
pending on SPM bit X72.
[0…20mA] Outputs 0…20mA.
[Range]
[4…20mA] Outputs 4…20mA.
[Hold] If an error occurs, the analog output retains
the last value.
[0mA] If an error occurs, the analog output is set to
0mA.
[4mA] If an error occurs, the analog output is set to
4mA.
[Output on wgt. err.]
[20mA] If an error occurs, the analog output is set to
20mA.
[linear] In the event of a negative weight, the analog
output is retained. This is only possible if the
output value for the zero weight is >0mA.
[0mA] In the event of a negative weight, the analog
output is set to 0mA.
[4mA] In the event of a negative weight, the analog
output is set to 4mA.
[On <0]
[20mA] In the event of a negative weight, the analog
output is set to 20mA.