Minebea Intec PR 5310 - Configuring the Analog Output

Minebea Intec PR 5310
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Operating | 6.2 - Configuration and adjustment via the keypad | 126
6.2.12 Configuring the analog output
Parameter table
Parameter Selection Description
[Off] Analog output is unused.
[Transparent] Value output in D30.
Instrument is controlled via PLC.
[Gross] Gross weight output in D8.
[Net] Net weight output in D9; only if tared, other
wise gross weight is output.
[Mode]
[Selected] Gross/net value output on the display (D11) de
pending on SPM bit X72.
[0…20mA] Outputs 0…20mA.
[Range]
[4…20mA] Outputs 4…20mA.
[Hold] If an error occurs, the analog output retains
the last value.
[0mA] If an error occurs, the analog output is set to
0mA.
[4mA] If an error occurs, the analog output is set to
4mA.
[Output on wgt. err.]
[20mA] If an error occurs, the analog output is set to
20mA.
[linear] In the event of a negative weight, the analog
output is retained. This is only possible if the
output value for the zero weight is >0mA.
[0mA] In the event of a negative weight, the analog
output is set to 0mA.
[4mA] In the event of a negative weight, the analog
output is set to 4mA.
[On <0]
[20mA] In the event of a negative weight, the analog
output is set to 20mA.