EasyManua.ls Logo

Minebea Intec PR 5310 - Testing the RS-232 Interface

Minebea Intec PR 5310
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Operating | 6.3 - Other functions via keypad | 141
6. Switch on the inputs externally.
The following appears on the display:
Digital input Identifier
Input 1 on 1 0 0
Input 2 on 0 1 0
Input 3 on 0 0 1
Input 1–3 on 1 1 1
Note: If the switched input is not displayed correctly, it is defective.
7. Press "Exit" to exit the test menu.
6.3.4 Testing the RS-232 interface
1. Press and hold for 2–3 seconds.
SEtuP appears on the display.
2. Press / .
tESt appears on the display.
3. Press "OK".
kEy appears on the display.
4. Press (3×).
rS232 appears on the display.
5. Press "OK".
1 1 appears on the display. The test starts automatically. The results
are displayed: -paSS- = OK, -FAIL- = error
6. Press "Exit" to exit the test menu.

Table of Contents

Related product manuals