10. SPM | 10.4 - System data | 171
10.4 System data
SPM address Data type R/W Function
X0–X2 BOOL R Digital input 1–3
X8–10 BOOL R Digital output 1–3
X16–18 BOOL R Output limit 1 1–3
B4
X32
X33
X34
X35
X36
X37
X38
X39
BYTE
BOOL
BOOL
BOOL
BOOL
BOOL
BOOL
BOOL
BOOL
R
R
R
R
R
R
R
R
R
Indicator status
ADC error
>Max (max. capacity; FSD = Full Scale Deflection)
>Max + permitted range (OVL)
<Zero
Zero ±¼d
Within the zeroset range (ZSR)
The weight is stable
Weight <Zero or >Max (max. capacity; FSD = Full Scale
Deflection)
B5
X40
X41
X42
X43
BYTE
BOOL
BOOL
BOOL
BOOL
R
R
R
R
R
ADC status
Measuring signal negative (Error7)
Measuring signal >36 mV (Error3)
Internal arithmetic error;
CAL data are perhaps faulty (Error1)
No or too low sense voltage (Error6)
B6
X48
X49
X50
BYTE
BOOL
BOOL
BOOL
R
R
R
R
Command status
Command error
Command active
Network failure signal
B7
X56
X57
X58
BYTE
BOOL
BOOL
BOOL
R
R
R
R
Active status
Test mode active
Calibration active
Device is tared