6. Operating | 6.2 - Configuration and adjustment via the keypad | 113
6.2.8 Limits
6.2.8.1 Functions of the limits
Each limit consists of a switch-on and a switch-off point for definition of a hys
teresis. The 3 pairs of values must be entered according to the same princi
ple. The limit values always refer to the gross weight. SPM addresses for the
limits, see ChapterSystem data.
Example 1:
The output signal (Limit1 out) of limit1 (Limit1) switches OFF above a
weight(Wgt) of 900kg.
The output signal (Limit2 out) of limit2 (Limit2) switches OFF below a weight
of 290kg.
The two limit values have a hysteresis of 10kg.
In the event of a power failure, the two outputs go to OFF, thus indicating un
der filling and over filling at the same time.