6. Operating | 6.2 - Configuration and adjustment via the keypad | 131
6.2.13 Entering operating parameters
1. Press and hold for 2–3 seconds.
SEtuP appears on the display.
2. Press "OK".
Cd 000 appears on the display (Cd flashes).
3. Select the individual parameters and modify if required, see Chapters
Parameter table (setup) and Selecting parameters.
[Address] oP 060 Addr
Enter device address, e.g. for printing.
Input: 01…99
[Sequence number] oP 061 SEqnr
The sequence number (counter for print jobs) is automatically
incremented (max. 999999) and can be set to a starting value here if
necessary.
The sequence number can also appear on the printout (selectable).
[Tare key] oP 062 tArkEy
The function of the tare key can be configured.
Selection: Set tare/reset tare, set tare/set tare again, disabled
[Set tare/reset tare] tAr.rES means that the device will be tared if it has
not been tared previously and the tare will be reset if the device has
already been tared.
[Set tare/ set tare again] tAr.tAr that upon every tare command, the
instant value in the tare memory is applied and the net display switches to
"0".
[Disabled] diSAbL means that the key has no function.
[Set zero key] oP 063 ZErkEy
The function of the zeroset key can be configured.
Selection: Only when not tared, reset tare on zero set, disabled
[Only when not tared] not.tAr limits the key function to gross mode.
[Reset tare on zero set] rES.tAr automatically switches the system to
gross mode.
[Disabled] diSAbl means that the key has no function.
[N.B.T key] oP 064 nbtkEy
The function of the key can be disabled diSAbl and enabled EnAbLE.
[Print key] oP 065 PrtkEy
The function of the key can be disabled diSAbl and enabled EnAbLE.
[Test key] oP 066 tStkEy
The function of the key can be disabled diSAbl and enabled EnAbLE.