EasyManua.ls Logo

Minebea Intec PR 5310 - 5.3.2 Electromagnetic Compatibility (EMC); RF Interference Suppression

Minebea Intec PR 5310
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15. Specification | 15.3 - Effect of ambient conditions | 184
15.3.2 Electromagnetic Compatibility (EMC)
All data in compliance with EN 61326 industrial section
High frequency electromagnetic
fields (80…2000MHz)
EN61000-4-3 10V/mHousing
Electrostatic discharge (ESD) EN61000-4-2 4/8kV
Fast transients (burst) EN61000-4-4 1kV
Peak voltages (surge) 1.2/50µs EN61000-4-5 1/2kV
Signal and control
lines
Conducted disturbances by
high frequency coupling (0.15…
80MHz)
EN61000-4-6 10V
Fast transients (burst) EN61000-4-4 2kV
Peak voltages (surge) 1.2/50µs EN61000-4-5 1/2kV
Conducted disturbances by
high frequency coupling (0.15…
80MHz)
EN61000-4-6 10V
Voltage dips EN 61000-4-11 40% / 0%
Mains inputs
Mains failure link EN 61000-4-11 20ms
15.3.3 RF interference suppression
Electromagnetic
emission
pursuant to EN61326, Limit classA, for
industrial areas