15. Specification | 15.3 - Effect of ambient conditions | 184
15.3.2 Electromagnetic Compatibility (EMC)
All data in compliance with EN 61326 industrial section
High frequency electromagnetic
fields (80…2000MHz)
EN61000-4-3 10V/mHousing
Electrostatic discharge (ESD) EN61000-4-2 4/8kV
Fast transients (burst) EN61000-4-4 1kV
Peak voltages (surge) 1.2/50µs EN61000-4-5 1/2kV
Signal and control
lines
Conducted disturbances by
high frequency coupling (0.15…
80MHz)
EN61000-4-6 10V
Fast transients (burst) EN61000-4-4 2kV
Peak voltages (surge) 1.2/50µs EN61000-4-5 1/2kV
Conducted disturbances by
high frequency coupling (0.15…
80MHz)
EN61000-4-6 10V
Voltage dips EN 61000-4-11 40% / 0%
Mains inputs
Mains failure link EN 61000-4-11 20ms
15.3.3 RF interference suppression
Electromagnetic
emission
pursuant to EN61326, Limit classA, for
industrial areas