5. Getting started | 85
5 Getting started
5.1 Switching on the device
The device is started up via the key pad of the indicator.
When the supply voltage is applied to the device, the following information is
displayed:
Pr5310
Device type PR 5310
01.00.00.
Firmware release
1350.50 kg
Weight display
Error6
Error message, if no sense voltage is connected,
see also ChapterError messages measuring cir
cuit.
Check the date and time after first turning on the indicator, see ChapterDate
& time.
5.2 Switching off the device
The device is switched off by disconnecting the supply voltage.
5.3 Device warm-up time
A warm-up time of30 minutes for the device is required before adjustment/
calibration is started.
5.4 Power failure
In the event of a power failure, all entered configuration, adjustment and
calibration parameters are retained.
5.5 Data backup
Additional overwrite protection is provided for adjustment/calibration data
and parameters (see ChapterCAL switch).