6. Operating | 6.2 - Configuration and adjustment via the keypad | 109
13. Press "OK".
The mV/V value for the max. capacity (Max) appears on the display.
14. Note the displayed value.
15. Press "Exit" to exit the setup menu.
6.2.7 Entering metrology parameters
1. Press and hold for 2–3 seconds.
SEtuP appears on the display.
2. Press "OK".
Cd 000 appears on the display (Cd flashes).
3. Press .
4. Select the individual parameters and modify if required, see Chapters
Parameter table (setup) and Selecting parameters.
[Measure time] CP 010 MEAtIM
The duration of a measurement can be selected.
Selection: 5ms, 10ms, 20ms, 40ms , 80ms, 160ms, 320ms, 640ms
[Digital filter] CP 011 FILtEr
Note: After changing the filter parameters, the maximum accuracy
should be reestablished through re-adjustment.