9. Fieldbus interface | 9.2 - Scale protocol | 158
Field Function
InsideZSR The gross weight value is within the zero setting range.
Standstill The scale is stable.
OutOfRange Below zero or above Max (FSD).
E9 The measuring signal is higher than the permissible range
of 36mV. Cannot read weight values from ADC (ana
log-digital converter) (Error 9).
E7 The measuring signal is negative (inverse conversion) (Er
ror 7)
E6 Sense voltage not present or too low (Error 6)
E3 The measuring signal is >36mV (no end of conversion) (Er
ror 3)
E1 Arithmetic error (overflow) (Error 1)
CmdError Error during execution (CmdError); e.g., the "taring" oper
ation is not processed, because the scale is not at a stand
still. The error is stored in LastError (function number 4).
The bit is reset with the ResetError bit (function num
ber121, see ChapterFunction number 112–124: transi
tion-controlled action bits (write)).
ActionActive The device is busy executing a function (e.g., waiting for
downtime for taring).
PowerFail Power failure; is always set after power on. The PowerFail
bit is reset with the ResetPWF bit (function number 85, see
ChapterFunction number 80–92: state-controlled action
bits (write)) "Reset power failure".
Test_Active The device executes the ADC test. The read weight value is
not the gross value, but the test value.
Cal_Changed The device has been calibrated.
When this bit is 1, the weighing parameters (EXPO/UNIT/
STEP) must be read again.
Set after "Power on" and reset after reading Max (FSD =
Full scale deflection).
Tare_Active The scale has been tared.