9. Fieldbus interface | 9.2 - Scale protocol | 167
Output range (MISO)
Byte Value Description
0 00 Gross weight - byte 0 (MSB)
1 00 Gross weight - byte 1
2 04 Gross weight - byte 2
3 D2 Gross weight - byte 3 (LSB)
4 08 Gross weight request de
tected.
5
6 Test
Ac
tive
7 Stand
still
In
side
ZSR
Cen
ter
Zero
Be
low
Zero
Over
load
Above
Max
ADC
Error
In direct access, status bits
are independent of the
write or read request.
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
The gross value (hex:000004D2 <=> 1234) can be read from bytes 0–3.
Negative values are output in the second complement.