11. Error messages | 11.2 - General error messages | 176
Display Error and Possible Cause
Err 71 A weight error occurs during calibration via the front panel if
the connection to the weighing point is interrupted.
Err 73 Dead load <-0.1 mV/V
Err 74 Arithmetic overflow
Err 75 Current weight < dead load
Err 76 Value for "d" is too high.
Err 77 Value for "d" is too low.
Err 78 Max < calibration weight.
Err 79 The weight cannot be displayed with the desired unit.
Err 80 Dead load + Max is too high.
Err 81 Max and scale interval do not match (1, 2, 5 10, 20, or 50).
Err 82 Not sufficient µV/d for W&M (legal-for-trade).
Err 83 Digital filter: Measurement time >160 ms.
Err 84 Filter frequency: Measurement time <40 ms/F
cut
.
Err 85 Filter frequency F
cut
is >40.00 Hz
Err 90 No printer interface assigned.