4. Device installation | 4.3 - Hardware construction | 40
4.3.3.3 Connecting a load cell with a 4-wire cable
Connection example panel device
Provide the following links between the terminal contacts:
from + Supply (+V) to + Sense (+S)
from – Supply (-V) to – Sense (-S)
Further connections:
Screen
Potential equalization
Connection example table-top device
Provide the following links between the terminal contacts:
from + Supply (+V) to + Sense (+S)
from – Supply (-V) to – Sense (-S)
Further connections:
Screen
Cable gland
Potential equalization