22
Procedures
2
Press the A/Z A key once.
The load display will show “CCAL”.
3
Press the “ENTER key. “CCAL mode can be
entered, then the load display will show “D01”
and it will light on and off.
When the calibration has completed already, the
set value of minimum scale which has registered
at that time will be displayed.
Set the minimum scale with A/Z OFF Y keys.
Setting value for the minimum scale will be 4
(four) as follows:
1, 2, 5, 10
A/Z OFF Y key Set value increment key
4
Press the “ENTER key.
The load display will show “DISP”.